1  Nuclei Are Tiny Magnets

1.1 In one sentence

What MRI looks at is the atomic nuclei inside the body (mainly hydrogen nuclei = protons). Some nuclei behave like tiny magnets (a property called spin), and when placed in a strong magnet they wobble around like a spinning top at a fixed rate. If you send in radio waves that exactly match this rotation rate (frequency), the nuclei respond. This is nuclear magnetic resonance (NMR) (Bloch 1946; Purcell et al. 1946).

1.2 Spin and the static field B0

Spin

A property of protons and other nuclei that is like “spinning on their own axis.” Because a charged object is rotating, it becomes a tiny magnet (a magnetic moment).

Static field B0

The constant, strong magnetic field produced by the MRI scanner’s large magnet. Its unit is the tesla (T). Clinical scanners use 1.5 T and 3 T; research scanners go to 7 T and beyond. That is tens of thousands of times the Earth’s magnetic field (about 0.00005 T).

Precession

The motion in which a spin in B0 wobbles around an axis along the direction of B0, just like the wobble of a tilted spinning top.

Magnetization M

The body contains an enormous number of protons. Even though each one points in a random direction, slightly more of them line up along B0, so overall a small magnet along B0 is formed (the longitudinal magnetization Mz).

1.3 The Larmor frequency

The precession frequency is proportional to the field strength.

\[ f_0 = \frac{\gamma}{2\pi} \, B_0 \]

  • \(f_0\): Larmor frequency (Hz)
  • \(B_0\): static field strength (T)
  • \(\gamma / 2\pi\): the gyromagnetic ratio, a constant fixed for each type of nucleus

For example, hydrogen (1H) has \(\gamma/2\pi \approx 42.58\) MHz/T, so on a 3 T scanner it precesses at about 128 MHz — a little higher than FM radio (76–95 MHz in Japan). The radio waves (RF) used in MRI are radio waves at this Larmor frequency.

TipKey point

Different nuclei have different \(\gamma\), so they precess at different frequencies even in the same scanner. That is why measuring 23Na or 31P requires a dedicated coil (antenna) tuned to that frequency (Chapter 6).

1.4 Main nuclei

Nucleus Spin γ/2π (MHz/T) Natural abundance Sensitivity relative to 1H What is commonly seen in the body
1H 1/2 42.58 99.99% 1 Water, fat
23Na 3/2 11.26 100% 0.093 Extracellular fluid, tissue Na
31P 1/2 17.24 100% 0.066 PCr, ATP, inorganic phosphate
13C 1/2 10.71 1.1% 0.016 Glycogen, metabolism (hyperpolarized)
19F 1/2 40.08 100% 0.83 Fluorinated drugs, contrast agents
39K 3/2 1.99 93.3% 0.0005 Intracellular K (ultra-high field)

Relative sensitivity is a rough guide to “signal strength for the same number of nuclei.” The actual signal is this multiplied by the concentration in the body.

Water hydrogen in the body is about 110 mol/L (water 55 mol/L × 2), whereas tissue sodium is only a few tens of mmol/L. Multiplying the relative sensitivity of 0.093 by this concentration difference, the 23Na signal is about 1/10,000 of the 1H signal (Madelin and Regatte 2013). This is why 23Na MRI images are coarse and take a long time to acquire.

1.5 Demo: how does the frequency change when you change the field?

1.6 Python: build a table of frequencies

code/01_larmor.py
# Compute the Larmor frequency f0 = (γ/2π) × B0 for each nucleus and field strength
gamma_bar = {  # γ/2π [MHz/T]
    "1H": 42.577,
    "19F": 40.078,
    "31P": 17.235,
    "23Na": 11.262,
    "13C": 10.708,
    "39K": 1.989,
}
fields = [1.5, 3.0, 7.0]  # [T]

print("nucleus " + "".join(f"{b:>9.1f}T" for b in fields))
for name, g in gamma_bar.items():
    print(f"{name:<7} " + "".join(f"{g * b:>8.1f}MHz" for b in fields))
nucleus       1.5T      3.0T      7.0T
1H          63.9MHz   127.7MHz   298.0MHz
19F         60.1MHz   120.2MHz   280.5MHz
31P         25.9MHz    51.7MHz   120.6MHz
23Na        16.9MHz    33.8MHz    78.8MHz
13C         16.1MHz    32.1MHz    75.0MHz
39K          3.0MHz     6.0MHz    13.9MHz

On a 3 T scanner, 23Na is at about 33.8 MHz. The 23Na coil actually used in a skeletal muscle 23Na study also operates at 32.6 MHz (around 3 T) (Marton et al. 2026).

1.7 Summary

  • A nuclear spin is a tiny magnet. In B0 it wobbles (precesses) at the Larmor frequency
  • The frequency is B0 × the gyromagnetic ratio γ. It differs by nucleus, so each nucleus needs its own coil
  • 23Na and 31P give far weaker signals than 1H (sensitivity × concentration)