1 Nuclei Are Tiny Magnets
1.1 In one sentence
What MRI looks at is the atomic nuclei inside the body (mainly hydrogen nuclei = protons). Some nuclei behave like tiny magnets (a property called spin), and when placed in a strong magnet they wobble around like a spinning top at a fixed rate. If you send in radio waves that exactly match this rotation rate (frequency), the nuclei respond. This is nuclear magnetic resonance (NMR) (Bloch 1946; Purcell et al. 1946).
1.2 Spin and the static field B0
Spin
A property of protons and other nuclei that is like “spinning on their own axis.” Because a charged object is rotating, it becomes a tiny magnet (a magnetic moment).
Static field B0
The constant, strong magnetic field produced by the MRI scanner’s large magnet. Its unit is the tesla (T). Clinical scanners use 1.5 T and 3 T; research scanners go to 7 T and beyond. That is tens of thousands of times the Earth’s magnetic field (about 0.00005 T).
Precession
The motion in which a spin in B0 wobbles around an axis along the direction of B0, just like the wobble of a tilted spinning top.
Magnetization M
The body contains an enormous number of protons. Even though each one points in a random direction, slightly more of them line up along B0, so overall a small magnet along B0 is formed (the longitudinal magnetization Mz).
1.3 The Larmor frequency
The precession frequency is proportional to the field strength.
\[ f_0 = \frac{\gamma}{2\pi} \, B_0 \]
- \(f_0\): Larmor frequency (Hz)
- \(B_0\): static field strength (T)
- \(\gamma / 2\pi\): the gyromagnetic ratio, a constant fixed for each type of nucleus
For example, hydrogen (1H) has \(\gamma/2\pi \approx 42.58\) MHz/T, so on a 3 T scanner it precesses at about 128 MHz — a little higher than FM radio (76–95 MHz in Japan). The radio waves (RF) used in MRI are radio waves at this Larmor frequency.
TipKey point
Different nuclei have different \(\gamma\), so they precess at different frequencies even in the same scanner. That is why measuring 23Na or 31P requires a dedicated coil (antenna) tuned to that frequency (Chapter 6).
1.4 Main nuclei
| Nucleus | Spin | γ/2π (MHz/T) | Natural abundance | Sensitivity relative to 1H | What is commonly seen in the body |
|---|---|---|---|---|---|
| 1H | 1/2 | 42.58 | 99.99% | 1 | Water, fat |
| 23Na | 3/2 | 11.26 | 100% | 0.093 | Extracellular fluid, tissue Na |
| 31P | 1/2 | 17.24 | 100% | 0.066 | PCr, ATP, inorganic phosphate |
| 13C | 1/2 | 10.71 | 1.1% | 0.016 | Glycogen, metabolism (hyperpolarized) |
| 19F | 1/2 | 40.08 | 100% | 0.83 | Fluorinated drugs, contrast agents |
| 39K | 3/2 | 1.99 | 93.3% | 0.0005 | Intracellular K (ultra-high field) |
Relative sensitivity is a rough guide to “signal strength for the same number of nuclei.” The actual signal is this multiplied by the concentration in the body.
Water hydrogen in the body is about 110 mol/L (water 55 mol/L × 2), whereas tissue sodium is only a few tens of mmol/L. Multiplying the relative sensitivity of 0.093 by this concentration difference, the 23Na signal is about 1/10,000 of the 1H signal (Madelin and Regatte 2013). This is why 23Na MRI images are coarse and take a long time to acquire.
1.5 Demo: how does the frequency change when you change the field?
1.6 Python: build a table of frequencies
code/01_larmor.py
# Compute the Larmor frequency f0 = (γ/2π) × B0 for each nucleus and field strength
gamma_bar = { # γ/2π [MHz/T]
"1H": 42.577,
"19F": 40.078,
"31P": 17.235,
"23Na": 11.262,
"13C": 10.708,
"39K": 1.989,
}
fields = [1.5, 3.0, 7.0] # [T]
print("nucleus " + "".join(f"{b:>9.1f}T" for b in fields))
for name, g in gamma_bar.items():
print(f"{name:<7} " + "".join(f"{g * b:>8.1f}MHz" for b in fields))nucleus 1.5T 3.0T 7.0T
1H 63.9MHz 127.7MHz 298.0MHz
19F 60.1MHz 120.2MHz 280.5MHz
31P 25.9MHz 51.7MHz 120.6MHz
23Na 16.9MHz 33.8MHz 78.8MHz
13C 16.1MHz 32.1MHz 75.0MHz
39K 3.0MHz 6.0MHz 13.9MHz
On a 3 T scanner, 23Na is at about 33.8 MHz. The 23Na coil actually used in a skeletal muscle 23Na study also operates at 32.6 MHz (around 3 T) (Marton et al. 2026).
1.7 Summary
- A nuclear spin is a tiny magnet. In B0 it wobbles (precesses) at the Larmor frequency
- The frequency is B0 × the gyromagnetic ratio γ. It differs by nucleus, so each nucleus needs its own coil
- 23Na and 31P give far weaker signals than 1H (sensitivity × concentration)