3  Photoplethysmography (PPG) and SpO₂

Both the heart rate on a smartwatch and SpO₂ on a pulse oximeter are measured with photoplethysmography (PPG). PPG is a method that shines light on the skin and captures pulsatile changes in blood volume from changes in the amount of light returning (or transmitted).

3.1 Principle: the Beer–Lambert law

When light passes through a substance, the amount absorbed is proportional to the concentration of the substance and the distance the light travels (the Beer–Lambert law).

I = I_0 \, e^{-\varepsilon(\lambda)\, c\, d}

  • I_0: incident light intensity, I: transmitted light intensity
  • \varepsilon(\lambda): absorption coefficient at wavelength \lambda, c: concentration, d: optical path length

Light passing through tissue is absorbed by skin, bone, venous blood, interstitial fluid, and so on, but these hardly change from beat to beat. On the other hand, the volume of arterial blood rises and falls with each heartbeat, so the optical path length d changes periodically.

flowchart TB
  subgraph Signal["Received light"]
    DC["DC component (large)<br/>tissue, venous blood,<br/>non-pulsatile arterial blood"]
    AC["AC component (small, 0.1 to a few % of DC)<br/>pulsatile arterial blood"]
  end
  Heart["Cardiac ejection"] --> Art["Rise and fall of arterial blood volume"] --> AC

The PPG signal is “a small AC component riding on a large DC component”. The AC/DC ratio (corresponding to the perfusion index) becomes small when peripheral blood flow is low, weakening the signal.

NoteReceived light and blood volume move in opposite directions

When blood increases, absorption increases and the received light decreases. Many devices invert the signal for display to make it easier to read.

3.2 Transmissive and reflective types

Transmissive Reflective
Arrangement LED and photodetector (PD) on opposite sides of the tissue LED and PD on the same side
Site Fingertip, earlobe Wrist, forehead, chest, finger (ring)
Signal Large and stable Small, easily affected by movement and pressure
Use Medical pulse oximeters Smartwatches, ring-type devices

3.3 Choosing the wavelength

Wavelength Characteristics Main use
Green (about 530 nm) Strong hemoglobin absorption; reaches only shallow layers → less affected by movement Heart rate measurement at the wrist
Red (about 660 nm) Deoxyhemoglobin (Hb) absorbs more than oxyhemoglobin (HbO₂) SpO₂
Infrared (about 940 nm) HbO₂ absorbs more than Hb. Reaches deep into tissue SpO₂, transmissive pulse wave

3.4 Principle of SpO₂: the ratio of ratios

Oxyhemoglobin and deoxyhemoglobin absorb red and infrared light in opposite ways. So we take the ratio of the AC/DC ratios at the two wavelengths, red and infrared.

R = \frac{(\mathrm{AC}/\mathrm{DC})_{\mathrm{red}}}{(\mathrm{AC}/\mathrm{DC})_{\mathrm{IR}}}

  • Dividing by DC (AC/DC) cancels out differences due to the individual and how the sensor is worn, such as LED brightness, skin color, and tissue thickness
  • Using only the AC component extracts information about only the pulsating arterial blood (the origin of the name “pulse” oximetry)

The lower the oxygen saturation, the relatively larger the red absorption (AC), and the larger R becomes. The conversion from R to SpO₂ is done not with a theoretical formula but with an empirical formula created from calibration data obtained by having healthy volunteers breathe hypoxic gas and sampling arterial blood. A commonly used textbook approximation is the following straight line.

\mathrm{SpO_2} \approx 110 - 25R\ [\%]

TipTry it
  • Halve the infrared AC/DC (e.g., the finger got cold and peripheral blood flow decreased). Confirm that if red decreases by the same proportion, SpO₂ does not change. This is why we take a ratio
  • In practice, when AC/DC is very small, the proportion of noise increases and the estimate of R becomes unstable

3.5 Limitations and caveats of PPG

3.5.1 Body movement

With wrist PPG, arm movement changes the blood in the tissue and the contact between skin and sensor, causing signal changes as large as or larger than the pulsations.

  • The arm-swing period during walking or running can be close to the heartbeat period and be mistaken for the heartbeat (cadence lock)
  • Many devices use adaptive filters or algorithms that remove the motion component by referring to the accelerometer signal

3.5.2 Low perfusion

When peripheral blood flow decreases due to cold, shock, vasoconstrictor use, etc., the AC component becomes small and the accuracy of heart rate and SpO₂ declines.

3.5.3 Skin color

Melanin absorbs light especially at shorter wavelengths (green, red).

  • A large retrospective study in the US (Sjoding et al. 2020) reported that “occult hypoxemia”, in which the pulse oximeter shows 92–96% although arterial oxygen saturation is below 88%, was about 3 times more frequent in Black patients than in White patients
  • Efforts are under way to increase the diversity of subjects from whom calibration data are collected and to revise how devices are evaluated

3.5.4 Others

  • Carbon monoxide poisoning: carboxyhemoglobin (COHb) absorbs red light similarly to HbO₂ and cannot be distinguished with two wavelengths → SpO₂ is displayed falsely high
  • Nail polish / artificial nails, ambient light, sensor pressure (too tight makes veins pulsate; too loose lets external light in)

3.6 Other information obtainable from PPG

  • Pulse rate, pulse rate variability (PRV): a surrogate for HRV, but not identical (HRV lecture material, Chapter 4)
  • Respiratory rate: estimated from respiratory variations in baseline, amplitude, and period
  • Pulse transit time (PTT): the time from the ECG R wave until the pulse wave reaches the periphery. When blood pressure rises, arteries stiffen and PTT shortens, so it is used for cuffless blood pressure estimation. However, individual calibration is required, the influence of vascular condition and posture is large, and accuracy is still at the research stage
  • Acceleration plethysmogram (second-derivative waveform): studied as an index of vascular stiffness

3.7 Summary

  • PPG = a large DC component + a small AC component (pulsatile arterial blood). The principle is the Beer–Lambert law
  • Green for heart rate; the two wavelengths red and infrared for SpO₂
  • SpO₂ is obtained from the ratio of ratios R using an empirical calibration formula
  • Accuracy declines with body movement, low perfusion, skin color, and carbon monoxide