flowchart LR SA["Sinus node<br/>(right atrium)"] --> Atria["Atrial muscle<br/>→ P wave"] Atria --> AV["AV node<br/>(slow conduction)<br/>→ PR interval"] AV --> His["Bundle of His"] His --> BB["Right and left bundle branches"] BB --> Pk["Purkinje fibers<br/>(fast conduction)"] Pk --> Vent["Ventricular muscle<br/>→ QRS complex"] Vent --> Rep["Ventricular repolarization<br/>→ T wave"]
1 Cardiac Electrophysiology and the ECG
1.1 The action potential of cardiomyocytes
Each heartbeat begins with an action potential, a sudden change in the membrane potential of cardiomyocytes. The shape of the action potential differs greatly between cell types.
| Ventricular muscle (working myocardium) | Sinus node (pacemaker cells) | |
|---|---|---|
| Resting membrane potential | Stable at about −90 mV | Not stable (depolarizes spontaneously from about −60 mV) |
| Upstroke (phase 0) | Steep, via Na⁺ current | Gradual, via L-type Ca²⁺ current |
| Plateau (phase 2) | Ca²⁺ influx and K⁺ efflux balance each other, lasting about 200–300 ms | No distinct plateau |
| Role | Contraction | Generating the rhythm |
The long plateau of ventricular muscle (= long refractory period) is the mechanism that prevents the myocardium from going into tetanus (sustained contraction).
1.2 The conduction system
- The AV node conducts slowly, creating a time delay so that the ventricles contract after atrial contraction has finished
- Purkinje fibers conduct very fast and excite the entire ventricle almost simultaneously → the QRS complex is short and sharp
1.3 ECG waveforms
| Wave / interval | Corresponding electrical activity | Typical normal values (adults) |
|---|---|---|
| P wave | Atrial depolarization | Width < 0.12 s |
| PR interval | Onset of atrial excitation to onset of ventricular excitation (mainly the AV-node delay) | 0.12–0.20 s |
| QRS complex | Ventricular depolarization | Width < 0.10–0.12 s |
| ST segment | Period when the whole ventricle is depolarized (plateau) | About the same level as the baseline |
| T wave | Ventricular repolarization | |
| QT interval | Ventricular depolarization to end of repolarization | Assessed after correction for heart rate |
Atrial repolarization is hidden in the QRS complex and is usually not visible.
1.3.1 Heart-rate correction of the QT interval
Because the QT interval shortens when the heart rate is fast, it is assessed as the corrected QT (QTc). A representative formula is Bazett’s.
\mathrm{QTc}_{\mathrm{Bazett}} = \frac{\mathrm{QT}}{\sqrt{\mathrm{RR}}} \qquad (\mathrm{QT},\ \mathrm{RR}\ \text{in seconds})
Because Bazett’s formula overcorrects at fast heart rates, Fridericia’s formula (\mathrm{QT}/\sqrt[3]{\mathrm{RR}}) and others are also used.
1.4 Why can the ECG be measured on the body surface? Dipoles and leads
1.4.1 The heart behaves like a “battery”
At the boundary between excited regions (extracellular side negative) and not-yet-excited regions (extracellular side positive), an electric dipole forms. The sum of the dipoles over the whole heart is called the cardiac vector, and its direction and magnitude change moment by moment during a beat.
Because the body is a volume conductor that conducts electricity, the electric field created by the cardiac vector reaches the body surface. Measuring the potential difference between two points on the body surface gives the component of the cardiac vector projected onto the direction connecting those two points. This “direction from which we look” is a lead.
When excitation (depolarization) travels toward the positive electrode, the ECG shows an upward deflection. When it travels away, the deflection is downward.
1.4.2 Standard limb leads (Einthoven’s triangle)
\begin{aligned} \mathrm{I} &= \Phi_{\mathrm{LA}} - \Phi_{\mathrm{RA}} \\ \mathrm{II} &= \Phi_{\mathrm{LL}} - \Phi_{\mathrm{RA}} \\ \mathrm{III} &= \Phi_{\mathrm{LL}} - \Phi_{\mathrm{LA}} \end{aligned} \qquad\Rightarrow\qquad \mathrm{II} = \mathrm{I} + \mathrm{III}
(RA: right arm, LA: left arm, LL: left leg potentials)
The 12-lead ECG consists of these 3 bipolar limb leads, 3 augmented unipolar limb leads (aVR, aVL, aVF), and 6 chest (precordial) leads (V1–V6). The limb leads view the heart in the frontal plane, and the chest leads in the horizontal plane.
1.4.3 Demo: cardiac vector and leads
Change the direction of the cardiac vector (angle in the frontal plane) and see how the amplitude of each lead changes. An angle of 0° points to the left (toward the patient’s left arm), and +90° points downward (toward the feet).
- The normal cardiac vector (electrical axis) is roughly −30° to +90°. Confirm that around 60°, the QRS in lead II is largest and aVR is most deeply negative
- Think about why, when detecting R waves for HRV analysis or heart-rate measurement, you choose a lead in which the R wave appears large
1.5 Properties of the ECG from a measurement perspective
| Property | Typical value | Implication for measurement |
|---|---|---|
| Amplitude | About 0.5–3 mV for the R wave | Hundreds of times smaller than the electrode DC offset (up to ±300 mV) → Chapter 2 |
| Frequency content | Mainly 0.05–100 Hz (QRS energy concentrated around 10–25 Hz) | Basis for filter design → Chapter 5 |
| Fastest change | QRS complex (tens of ms) | Determines the lower limit of the sampling frequency → Chapter 4 |
1.6 Summary
- Sinus node → atria (P wave) → AV node (PR) → bundle of His, Purkinje fibers → ventricles (QRS) → repolarization (T wave)
- The ECG is the cardiac vector projected onto the direction of each lead. The waveform changes with the lead
- The ECG is a small signal on the order of mV, with components of about 0.05–100 Hz